User's Guide

28
Specifications
Item
Subcomponent
Transmitter
Sensor
PDA
Model Number
G7-T01
G7-T01A
G7-T01B
G7-S01
G7-S01A
G7-S01B
G7-P02
Operating
Temperature
5-40℃(41-104℉)
Operating
Humidity
10-93%non-condensing
Storage and
Transportation
Temperature
-20-60
4-30
-20-60
Storage and
Transportation
Humidity
5-95%non-condensing
Storage and
Transportation
Pressure
700hpa~1060hpa
Ingress
Protection
Level
IPX7
IPX0
Use Life
4 Years
G7-S01A14 days
G7-S01B10 days
G7-S01C7 days
Shelf life: 1 year
4 Years
Detection
Range
2.0mmol/L-25.0 mmol/L
Measurement
accuracy
When the glucose concentration >4.2mmol/l (75mg/dL), the
accuracy deviation of the sensor does not exceed ±20%; when
the glucose concentration 4.2 (75mg/dL), the accuracy
deviation does not exceed ±1mmol/l (18mg/dL).
Wireless
Frequency
and
Bandwidth
Frequency2.402GHz ~ 2.48 GHz
Bandwidth1Mbps
Wireless
Modulation
GFSK