User's Guide
29 
【Electromagnetic Compatibility】 
These devices are intended for use in the electromagnetic environment specified below. The 
customer or the user of the device should assure that the device is used in such an environment. 
Portable and mobile RF communication interference may have an impact on the device. 
Please use the cables and accessories provided. The cable information is as follows: 
# 
Item 
Length(m) 
Shielded? 
Notes 
1 
PDA CHARGING CABLE 
1.0m 
YES 
EUT      DC 5V 
The use of accessories other thanthose specified for the device is not recommended. They 
may result inincreased emissions or decreased immunity of the device. 
The device should not be used adjacent to or stacked with other equipment. If adjacent or 
stacked  use  is  necessary,  the  device  should  be  observed  to  verify  normal  operation  in  the 
configuration in which it will be used. 
The basic performance is described in the table below: 
Performance   
Description 
Measurement 
Accuracy 
When the glucose concentration is >75mg/dL, the sensor 
accuracy deviation does not exceed ±20%; 
When the glucose concentration is ≤75mg/dL, the accuracy 
deviation does not exceed ±20mg/dL 
IEC60601-1-2 Table 201 
Guidance and manufacturer's declaration  –  electromagnetic immunity 
The device is intended for use in the electromagnetic environment specified below. The customer 
or the user of the device should ensure that it is used in such an environment. 
Emissions Test 
Compliance 
Electromagnetic environment - guidance 
RF Emissions CISPR 11 
Group 1 
The  device  uses  RF  energy  only  for  its  internal 
function. Therefore, its RF emissions are very low and 
are  not  likely  to  cause  any  interference  in  nearby 
electronic equipment. 
RF Emissions CISPR 11 
Class B 
The  device  is  suitable  for  use  in  all  establishments, 
including  domestic  establishments  and  those  directly 
connected to the public low-voltage power supply. 
Harmonic Emissions IEC   
61000-3-2 
Class A 
Voltage 
Fluctuations/Flicker 
Emissions   
IEC 61000-3-3 
Complies 
IEC 60601-1-2:Table 202 
Guidance and manufacturer’s declaration - electromagnetic immunity 
The device is intended for use in the electromagnetic environment specified below. The customer 
or the user of the device should ensure that it is used in such an environment. 
Immunity Test 
IEC 60601 Test Level 
Compliance Level 
Electromagnetic environment - 
guidance 