User's Guide
31 
Radiated RF 
IEC 61000-4-3 
150kHz~80MHz 
10V/m 
80MHz~2.5GHz 
10V/m 
the transmitter. 
Recommended separation 
distance 
d=1.2  √P 
d=1.2  √P 80MHz~800MHz 
d=2.3  √P 800MHz~2.5GHz 
where P is the maximum output 
power rating of the 
transmitter in watts (W) 
according to the transmitter 
manufacturer and d is the 
recommended separation 
distance in metres (m).Field 
strengths from fixed RF 
transmitters, as determined by 
an electromagnetic 
site survey,a should be less than 
the compliance level 
in each frequency range.b 
Interference may occur in the 
vicinity of equipment 
marked with the following 
symbol:  
IEC60601-1-2: Table 206 
Recommended separation distances between portable and mobile RF 
communications equipment and the device 
These devices are intended for use in an environment in which radiated RF disturbances are 
controlled. The customer or the user of the device can help prevent electromagnetic interference 
by maintaining a minimum distance between portable and mobile RF communications 
equipment(transmitters) and the device as recommended below, according to the maximum 
output power of thecommunications equipment 
Maximum rated 
output power of 
Transmitter (W) 
Separation distance according to frequency of transmitter (m) 
150kHz~80MHz 
d=1.2√P 
80MHz~800MHz 
d=1.2√P 
800MHz~2.5GHz 
d=2.3√P 
0.01 
0.12 
0.12 
0.23 
0.1 
0.38 
0.38 
0.73 
1 
1.2 
1.2 
2.3 
10 
3.8 
3.8 
7.3 
100 
12 
12 
23 
For transmitters rated at a maximum output power not listed above, the 