Technical Specifications
82100-0043 Issue 2
THURLBY THANDAR INSTRUMENTS
TSX Series
High Power Laboratory DC Power Supplies
350/360 watts, standard and bus programmable
OUTPUT SPECIFICATIONS
Operating modes: Constant voltage or constant current with
automatic crossover.
Voltage range: 0V to 35V (TSX3510/TSX3510P).
0V to 18V (TSX1820/TSX1820P).
Current range: 0A to 10A (TSX3510/TSX3510P).
0A to 20A (TSX1820/TSX1820P).
Overvoltage protection: 10% to 110% of max. output voltage.
Setting resolution: 10mV, 10mA.
Load regulation: <0.01% of max. O/P for 90% change.
Line regulation: <0.01% of max. O/P for 10% change.
Output impedance:
<1min constant voltage mode.
>5kin constant current mode.
Ripple & noise: <1mV RMS typical in constant voltage.
<3mA RMS typical in constant current.
HF common mode noise: Typically <3mV RMS, <10mV pk.
Transient load response: <20us to within 50mV of set level for
90% load change.
Temperature coefficient: typically <100ppm/
o
C.
Overvoltage protection
delay:
<200us.
Protection functions: Overvoltage trip
Regulator overtemperature
Sense miswiring.
Status indication: Output On lamp
Constant voltage mode lamp
Constant current mode lamp
Trip message on display.
Output terminals: 4mm output terminals at front, screw ter-
minals for output and sense at rear.
Output protection: Full forward and reverse protection via
OVP and diode clamp.
METER SPECIFICATIONS
Meter types: Separate 4 digit meters for voltage and
current with 12.5mm (0.5") LED displays.
Meter resolutions: 10mV, 10mA.
Meter accuracies: Voltage ±(0.2% + 1 digit)
Current ±(0.5% + 1 digit).
GENERAL
Input voltage range: 220V to 240 volts ± 10% or 110 to 120
volts ± 10%, 47 to 63Hz.
Installation Category II.
Power requirement: 600VA max.
Electrical safety: Complies with EN61010-1.
EMC: Complies with EN61326.
Temperature: +5
o
C to +40
o
C operating, 20% to 80%
RH, -40
o
C to +70
o
C storage.
Cooling: Silent fan-less convection cooling.
Size: 210 x 130 x 350mm (WxHxD)
(half rack width x 3U height).
Weight: 5.0kg (TSX versions).
5.5kg (TSX-P versions).
Options: Rack Mounting kit.
FRONT PANEL CONTROLS (standard versions)
Voltage setting: Via single rotary controls for coarse and
fine control.
Current setting: Via single turn semi-logarithmic rotary
control.
Overvoltage setting: Via screwdriver adjustable preset poten
-
tiometer.
Output On/Off: Latchinh push-push switch operating
electronic power control.
FRONT PANEL CONTROLS (P versions)
Voltage setting: Direct keyboard entry or quasi-analogue
rotary control.
Current setting: Direct keyboard entry or quasi-analogue
rotary control.
Overvoltage setting: Direct keyboard entry.
Output On/Off: Push button control with indicator lamp
operating electronic power control.
Note: all voltage and current levels set via the keyboard are displayed
on a separate 0.3" 4 digit display. This entry preview system ensures
that the user can observe the value entered before it is effected thus
avoiding possible error. The display is also used for setting additional
functions and for displaying watts.
Additional keyboard functions:
Increase or decrease voltage or current in user-selectable steps
(delta mode). Store/recall voltage, current & OVP levels from
non-volatile memory (25 memories). Set digital interface type
(RS232 or GPIB), set baud rate, set address.
DIGITAL INTERFACES (P versions)
RS232:
Variable baud rate, 9600 baud maximum, 9 pin D connector
(male). Fully compatible with standard RS232 or TTi address-
able RS232 system (ARC).
IEEE-488 (GPIB): Conforming with IEEE488.1 & IEEE488.2.
Bus Functions: Set voltage; set current; set OVP; set output
On/Off; read output voltage/current.
Setting resolution: Voltage - 10mV; Current - 10mA.
Setting accuracy: Voltage - ±(0.1% + 10mV);
Current - ±(0.2% + 20mA).
Response times:
Interface - <15ms (single command);
PSU - Depends on Load conditions, typically 150ms to within
0.1% of final value (except for voltage reduction with low load
current which will be longer).
Readback resolution: Voltage - 10mV; Current - 10mA.
Readback accuracy: Voltage - ±(0.1% + 1 digit);
Current - ±(0.5% +1 digit).
Operating software:
Software for operating the PSUs under GPIB or RS232 control
is available including a Labwindows* driver and ARC-TALK
software for a PC.
Models Voltage/Current Power
TSX1820 TSX1820P 18 volts, 20 amps 360 watts
TSX3510 TSX3510P 35 volts, 10 amps 350 watts
Designed and built in Europe by:
Thurlby Thandar Instruments Ltd. operates a policy of continuous development and
reserves the right to alter specifications without prior notice. * Labwindows is a trade
-
mark of National Instruments Corporation.
Thurlby Thandar Instruments Ltd.
Glebe Road, Huntingdon. Cambs. PE29 7DX United Kingdom (UK)
Tel: +44 (0)1480 412451 Fax: +44 (0)1480 450409
Email: sales@tti-test.com Web: http://www.tti-test.com