Companion 5TM Oxygen Concentrator User Manual
User Controls & System Status Indicators Symbol 02 Definition Symbol Read user manual before operation. See user manual for instructions. I/O Definition On/Off Switch No Smoking Icon: Do not smoke near unit. Use no oil or grease. Warnings / ALERT (Yellow) Indicator No open or naked flames. Class II equipment No serviceable parts inside. Do not open cover. Oxygen Output Certified for both the U.S. and Canadian markets, to the applicable U.S. and Canadian standards.
Contents User Controls & System Status Indicators...............................................................................................2 Quick Start Guide.......................................................................................................................................4 Indications for Use.....................................................................................................................................7 Contraindications for Use.................................
Quick Start Guide 1 Unpack Your Companion 5 Companion 5 2 Getting to Know Your Companion 5 Review all Warnings, Cautions and additional device information in the rest of this manual. Become familiar with the key features of the Companion 5 and the User Control Panel.
The Companion 5 Components Outer Case Power Switch Durable plastic case that encloses and protects the inner components of the CAIRE Companion 5. ON/OFF power switch used to initiate or stop power supply to the unit. The ON position is indicated by the “l” symbol, and the OFF position is indicated by the “O” symbol. Outlet Barb Oxygen exits the Companion 5 here after it has been filtered and concentrated. Either a single lumen nasal cannula or standard oxygen tubing is attached to deliver to the patient.
4 5 Power On and Warm Up Adjust Flow Setting to Prescribed Level Use the Flow Control Knob to select the flow prescribed by your physician. Plug the AC power cord into an AC outlet. Toggle the Power Switch to the On position. When the Companion 5 is plugged in properly and the switch is toggled into the “ON” (I) position, the external power indicator will appear on the user control panel. Power Switch When the Companion 5 is plugged in properly, a green indicator on the LED Display will light up.
Important! Safety Instructions are defined as follows: WARNING: IMPORTANT SAFETY INFORMATION FOR HAZARDS THAT MIGHT CAUSE SERIOUS INJURY. CAUTION: Important information for preventing damage to the Companion 5. Note: Information needing special attention. Indications for Use The Companion 5 is intended for the administration of supplemental oxygen. The device is not intended for life support, nor does it provide any patient monitoring capabilities.
WARNING: Operating the oxygen concentrator outside of the operational temperature specifications can limit the concentrator’s ability to meet Oxygen Concentration specification. Refer to the specification section of this manual for temperature limits. WARNING: Use no oil, grease, or petroleum-based or other flammable products with the oxygen-carrying accessories or the Oxygen Concentrator. Only water based, oxygen compatible lotions or salves should be used.
Recommended Operating Environments For proper use of the device, the following chart provides important information concerning the recommended operating environments, or operating conditions. Specifications Operating 41º F to 104º F (5° to 40°C) Temperature Operating Humidity 10% - 95% at an 82.4° F (28° C) dew point Transport/ -40º F to 158º F (-40º C to 70º C) Storage Humidity: 10% - 95% Non-Condensing Temperature Electrical Use no extension cords. Use no electrical outlets controlled by a switch.
Operating Instructions Before Operating Step 2: If the Bubble Humidifier WILL NOT be Used This user manual serves as your reference to help you operate and maintain the device. If you have any questions or concerns please call your home care provider. a. Connect a nasal cannula or oxygen tubing to the oxygen outlet barb. Important! DO NOT attempt to operate the Companion 5 without first reading the Safety Guidelines section of this manual. Please follow all of the operating instructions.
e. Connect one end of the humidifier tube to the lid of the humidifier bottle. Step 4: Power On and Warm Up a. Plug the electrical cord of the CAIRE Companion 5 into an AC outlet. WARNING: Inspect the electrical cord for damage before use. If the cord is damaged, do not plug it into an electrical outlet or attempt to operate the concentrator. f. Connect the other end of the humidifier tube to the outlet barb of the CAIRE Companion 5. b. Press the power switch in the “ON” (I) position.
Step 5: Adjust Flow Control Rate Step 6: Verify Flow and Breathe Normally a. Turn the flow control knob to the oxygen flow rate (LPM) prescribed by your physician. a. Verify that oxygen is indeed flowing through the nasal cannula and that there are no kinks, bends, or blockages in the tubing. Note: Ensure the cannula is fully inserted and secure. You should hear or feel oxygen flow to the prongs of the nasal cannula. b. Properly position your nasal cannula and breathe normally through the nasal cannula.
Alarm Conditions Step 7: Turn Off a. Press the power switch in the “OFF” (O) position when the CAIRE Companion 5 is no longer in use. b. Store the Companion 5 in a cool and dry location for a period no longer than 1 month. The Companion 5 must run for a minimum of 2 hours per month to ensure proper function of the unit. CAUTION: Use only accessories recommended by the manufacturer. Use of any other may be hazardous, cause serious damage to your oxygen concentrator and will void the warranty.
Alarm Conditions Table Audible Alarm Colored LED Off Green Solid Continuous RED Solid Continuous RED Solid Cause Your Action The concentrator is working properly. None The concentrator has lost power but the power switch is still in the “ON” position. Verify that the concentrator’s electrical cord is plugged into an outlet and that the outlet has power. Try a different outlet. If the problem continues, connect to a back-up oxygen supply and contact your healthcare provider.
Cleaning, Care & Routine Maintenance Routine Maintenance Power Cord and Outer Case: Servicing of the internal components inside the cabinet of the Companion 5 must be conducted by a CAIRE trained and qualified service technician as needed. Turn OFF the Companion 5 and disconnect from AC power before any cleaning or disinfecting activity. DO NOT spray the outer case directly. Use a damp (not wet) cloth or sponge. Spray the cloth or sponge with a mild detergent solution to clean the cabinet and power cord.
CAIRE Inc. 2200 Airport Industrial Drive, Ste. 500 Ball Ground, GA 30107 USA 1-800-482-2473 Worldwide: www.CAIREmedical.com Medical Product Services GmbH Borngasse 20 35619 Braunfels, Germany Worldwide: www.CAIREmedical.com CAIRE Inc. reserves the right to discontinue its products, or change the prices, materials, equipment, quality, descriptions, specifications and/or processes to its products at any time without prior notice and with no further obligation or consequence.