Recommended Operating Environments
For proper use of the device, the following chart provides
important information concerning the recommended operating
environments, or operating conditions.
Specifications
Operating
Temperature
50º F to 104º F (10º C to 40º C)
Operating
Humidity
10% - 95% at an 82.4° F (28° C) dew point
Transport/
Storage
Temperature
-4º F to 140º F (-20º C to 60º C)
Humidity: Up to 95% Non-Condensing
Electrical Use no extension cords. Use no electrical outlets con-
trolled by a switch.
Altitude 0 – 13,123 feet (0 - 4,000 meters)
Placement DO NOT block the air inlet (top back cover) or the exhaust
vent (back bottom left). Place device a minimum of 3 inch-
es (7.5 cm) away from walls, draperies, furniture, etc.
Environment Must be smoke, pollutant, and fume free.
Operating
Time
24-hours a day when connected to an external AC or DC
power source.
Proper Placement of the Eclipse
Select a location for the device that avoids the intake of smoke,
fumes and pollutants. Correct placement of the device should
allow intake of air through the air inlet lter at the top rear of the
cabinet and allow exhaust air to freely leave the exhaust vent at
the bottom left of the device.
Place the device such that the alarms can be heard.
Position the oxygen supply tubing in such a way that it does not
kink or occlude.
Note: If the Oxygen Concentrator has been stored for an extended period
of time outside its normal operating temperature range, the unit should
be allowed to return to normal operating temperature before being turned
on. (Refer to the Specifications section in this manual.)
Note: When the automobile in which you are using the Oxygen Con-
centrator unit is turned off, disconnect and remove the device from the
automobile. Do not store the Oxygen Concentrator in a very hot or cold
automobile or in other similar, high-or low-temperature environments.
DO NOT leave the Oxygen Concentrator or the Power Supply plugged into
the vehicle if the ignition is in the off position. Doing so may drain the
vehicle’s battery.
WARNING: DO NOT USE YOUR OXYGEN CONCENTRATOR IN
THE PRESENCE OF FLAMMABLE GASES. THIS CAN RESULT
IN RAPID BURNING CAUSING PROPERTY DAMAGE, BODILY
INJURIES OR DEATH. USE NO OIL, GREASE, OR PETROLEUM-BASED
OR OTHER FLAMMABLE PRODUCTS WITH THE OXYGEN-CARRYING
ACCESSORIES OR THE OXYGEN CONCENTRATOR. ONLY WATER
BASED, OXYGEN COMPATIBLE LOTIONS OR SALVES SHOULD BE USED.
OXYGEN ACCELERATES THE COMBUSTION OF FLAMMABLE SUB-
STANCES.
WARNING: THIS DEVICE SUPPLIES HIGH-CONCENTRATION OXYGEN
THAT PROMOTES RAPID BURNING. DO NOT ALLOW SMOKING OR
OPEN FLAMES WITHIN THE SAME ROOM OF (1) THIS DEVICE, OR (2)
ANY OXYGEN-CARRYING ACCESSORY. FAILURE TO OBSERVE THIS
WARNING CAN RESULT IN SEVERE FIRE, PROPERTY DAMAGE, AND/
OR CAUSE PHYSICAL INJURY OR DEATH.
CAUTION: Do not operate unit in a restricted or confined
space (i.e., a small case or handbag) where ventilation
can be limited. This can cause the Oxygen Concentrator to
overheat and impair performance. Do not allow either the air intake
or the air outlet vents to become blocked. This can cause the Oxygen
Concentrator to overheat and impair performance.