Product Manual

FreeStyle 5 EN
AirSep
®
Corporation EN-31
MN167-1 rev. 02/13
Alarm
Rapid
intermittent:
Beep, beep,
beep…
(Yellow) alarm;
intermittent light
Breathing rate is
exceeding the
capacity of the
FreeStyle 5 unit.
Reduce activity, then if
necessary use another
source of oxygen as
available. Contact
your Equipment
Provider.
Alarm
Rapid
intermittent:
Beep, beep,
beep…
(Red) alarm;
continuous light
General
malfunction of the
FreeStyle unit has
occurred.
Turn off the unit.
Change to another
source of oxygen, and
contact your
Equipment Provider.
Figure 16: Alarms Indicator
Dual display:
Alarm condition (yellow)
General malfunction (red)