Product specifications

MN078-1 Rev F
8/12
10
[Read the Important Safety Rules section before operating this equipment.]
6.
Insert power cord into the electrical outlet.
7.
Locate the power switch on the front
of the unit, and switch it to
the
I
position (on). (Figure 2.)
A battery
-operated audible alarm must sound for a 5-
second
test to indicate a good battery and alarm.
If the alarm is weak or does not sound at all, consult
your Equ
ipment Provider immediately.
Figure 2
The standard New
Life Elite
Oxygen
Concentrator
accommodates
prescriptions
from 1 to 5 lpm
.
8.
Dual Flow and High Flow Applications:
Set the flowmeter adjustment
knob(s) to the prescribed lpm, in any combination of flows up to a total
of 5 lpm for Dual 5 un
its (Figure
3).
or
Pediatric/Low Flow Applications:
Set the flowmeter adjustable knob(s)
to the prescribed lpm up to 2 lpm in 1/8 liter (125 ccm) increments.
Figure 3
The concentrator is now ready for use.
On/Off
Power Switch