11) Apply 5000, 10000, 15000, and 19000 lbs and verify span is within 1.0% of
applied load. Make any adjustments necessary to bring the unit into
tolerance. Insure that unit does not indicate overcapacity before 20000 lbs
of applied weight.
12) Apply Torx-seal to lock the potentiometer shaft.
13) Reassemble keyboard assembly.
14) Replace bezel screws.