When the instrument is set to autoacquire mode, the FAST
tracka
will automatically generate a
filename with a format corresponding to the acquisition date and time.
5.3 FAST
TRACKA
SYSTEM MENUS
5.3.1 System Setup Menu
System Setup has 14 options and displays the present setup parameters:
=====================
System Setup
=====================
Fast Repetition Rate Fluorometer - Ver 1.0
FPGA Version - Ver 0.1
Instrument ID - Ser
Flashcard Size - 20 MB
AutoAcquire is DISABLED
Tue June 10 10:36:48 1997
Controller Battery Voltage = 14.88 V
Controller Current = 0.090 A
Electronics Temp = 22.91 Deg C
A: Set Date and Time
B: Boot protocol slot number - 0
C: AutoAcquire is DISABLED
D: REF Amplifier offset (counts)- 128
E: PMT Amplifier offset (counts)- 120
F: Reserved
G: Reserved
H: F0 analog output scale maximum - 1.00000
I: FM analog output scale maximum - 1.00000
J: PMT calibration threshold is - 200 counts
K: Ref calibration threshold is - 200 counts
L: Set PMT gain constants
M: Check PMT calibration
X: Reset to Safe values
Select option or '0' to return:
SYSTEM SETUP Header:
Fast Repetition Rate Fluorometer - Ver 1.0
FPGA Version - Ver 0.1
Instrument ID - Ser
Flashcard Size - 20 MB
AutoAcquire is DISABLED