Universal precautions shall be observed by all people who come in contact with the
instrument and/or accessories to help prevent their exposure to blood-borne pathogens
and/or other potentially infectious materials. In any circumstance, wherein the exact
status of blood or body fluids/tissues encountered are unknown, it shall be uniformly
considered potentially infectious and handled in accordance with OSHA guidelines.
EMC Statement
It is important to install and use the equipment in accordance with the instructions
in order to prevent harmful interference with other devices in the vicinity. If this
equipment causes harmful interference to other devices (determined by turning the
equipment off and on), the user is encouraged to try to correct the interference by one
or more of the following measures:
Reorient or relocate the other device(s).
Increase the distance between the equipment.
Connect this equipment into an outlet on a circuit different from that to which the
other device(s) is connected.
Consult the manufacturer or your Alcon field service engineer for help.
Table 1-2 Guidance and Manufacturer's Declaration - Electromagnetic Emissions - The Constellation®
Vision System is intended for use in the electromagnetic environment specified below. The
customer or the user of the system should assure that it is used in such an environment.
Emissions Test Compliance Electromagnetic Environment-Guidance
RF emissions
CISPR 11
Group 1 The Constellation
®
Vision System uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause any interference
in nearby electronic equipment.
RF emissions
CISPR 11
Class A Based on extensive field experience the Constellation
®
Vision System is suitable
for use in all establishments other than domestic and those directly connected
to the public low voltage power supply network that supplies buildings used for
domestic purposes.
The EMC Statement provides guidance on steps to take in case of electromagnetic
interference.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
Flicker emissions
IEC 61000-3-3
Complies