User's Manual Part 1

1.18  8065751025
Constellation
®
 Vision System
Cautions and Warnings
Footswitch
Never pick up or move the footswitch by holding the cable. Damage may result.
Cassette
During initialization the drain pump is rotated to the home position; therefore, keep 
hands and fingers clear of cassette well during power-on initialization. Manually 
rotating the hub roller in the cassette well when power is on and a cassette is not 
installed can cause incorrect cassette loading and/or can cause injury to fingers.
WARNINGS!
All  fluids  aspirated  during  surgery  should  be  treated  as  biohazards.  Take 
appropriate  precautions  when  handling  instruments  and  lines  in  contact  with 
aspirated fluids.
Drain  bag  volume  should  not  exceed  500cc  “Max.  Capacity.”  Exceeding  this 
volume may result in a biohazardous condition.
Consumables
Do not use consumable paks beyond the expiration date stamped on the outer 
packaging. Sterile consumable medical devices should not be reused (Accreditation 
Manual for Hospitals, 1982); they are intended for single use only. Improper usage 
or assembly could result in a potential hazardous condition for the patient. Alcon 
assumes no responsibility for complications that may arise as a result of the reuse or 
improper usage of consumables.
The equipment used in conjunction with Alcon 
Constellation
®
 Vision System 
consumables constitutes a complete system. Use of consumables other than Alcon 
consumables may affect system performance and create potential hazards, and if it 
is determined to have contributed to the malfunction of the equipment under service 
contract, could result in the voidance of the contract and/or invoicing at prevailing 
hourly rates.
WARNING!
Attach only Alcon supplied consumables to console and cassette luer fittings. Do 
not connect consumables to the patient's intravenous connections.
In all cases, the instrument setup instructions contained in this manual, and all label 
instructions in the package, should be thoroughly understood prior to using any of the 
Constellation
®
 Vision System Pak configurations.
Ensure that tubing is not occluded during any phase of operation.