User's Manual Part 2

2.28 8065751025
Constellation
®
Vision System
Menu Bar: Options - SYSTEM SETTINGS
Selecting "System" from the Options menu displays the screen shown in Figure 2-17.
This popup screen allows the user to configure settings that are part of the system's
normal operation. Once saved by pressing the Close button, the settings become the
default when the system is powered up.
SYSTEM SETTINGS - SETTINGS
The settings screen allows adjustments to the following system parameters:
Screen Brightness - Sets the relative brightness of the touch screen. The
allowable range is 1 – 9 with 7 as the default.
Language - Allows the user set the language used for system purposes such as end
case reports.
WiFi Status - Allows the user determine whether or not WiFi functionality will be
available.
Patient Eye Level Offset - Allows the user set the difference in height between the
cassette and the patient’s eye level. Range: 0 – 100.
Patient Eye Level Offset Units - Allows the user determine the units (inches or
cm) associated with Patient Eye Level Offset values.
Auto Gas Fill Purge Cycles - Allows the user determine the number of purge
cycles (0-3) when performing an AGF.
Figure 2-17 System Settings - Settings Screen