User's Manual Part 2

2.38 8065751025
Constellation
®
Vision System
The More Information popup for Infusion allows the user to quickly set values for its
two different set points:
Regular infusion - Infusion value the system will go to upon entering infusion
mode.
Alternate infusion - Infusion value the system will go to when the footswitch
button is pressed that is assigned the "InfusionFaxAltRegularToggle" action.
Additionally, the popup contains a change bottle button that will open the Fluid
Container Setup popup discussed later in this section.
Fluid/Air Exchange (F/AX) Global Control
The F/AX global control turns F/AX on/off, adjusts setpoints and brings up advanced
settings. The More Information popup is design to enable the user to quickly set
values for its two setpoints:
Regular F/AX - F/AX value the system will go to upon entering F/AX mode.
Alternate F/AX - F/AX value the system will go to when the footswitch button is
pressed that is assigned the "InfusionFaxAltRegularToggle" action.
Figure 2-26 F/AX Global Control and Advanced Setup Popup
F/AX Set point/Actual Value
F/AX On/
Off Button
More
Information
Button
Increase Value Button
Decreases Value Button
F/AX Source
F/AX Units
Irrigation Global Control
The Irrigation global control turns Irrigation On/Off, adjusts setpoints and brings up
advanced settings. The advanced setup popup for Infusion allows the user to quickly
set values for its two different set points:
Regular Irrigation - Irrigation value the system will go to upon entering Irrigation
mode.
Alternate Irrigation - Irrigation value the system will go to when the footswitch
button is pressed that is assigned the "InfusionFaxAltRegularToggle" action.
Additionally, the popup contains a change bottle button that will open the Fluid