User's Manual Part 2

8065751025 2.39
Constellation
®
Vision System
Diathermy Global Control
Diathermy Global control turns diathermy on/off, enables fixed or proportional
mode, adjusts setpoints, and displays an advanced configuration screen that allows
diathermy values to be changed in more detail. Selecting Fixed diathermy provides
a fixed percentage of diathermy power when the footswitch treadle is pressed.
Proportional diathermy increases the diathermy power in proportion to the distance
the footswitch treadle is depressed.
Irrigation Set point/Actual Value
Irrigation
On/Off
Button
Advanced
Setup
button
Increase Value button
Decrease Value button
Irrigation Source
Irrigation Units
Figure 2-27 Irrigation Global Control and Advanced Setup Popup
Diathermy Set point/Actual Value
Mode Toggle Button
Advanced
Setup
Increase Value button
Decrease Value button
Fixed
Proportional
Figure 2-28 Diathermy Global Control and Advanced Setup Popup