SYSTEM STATE AREA
The System State area provides specific information on one of the following available
states : Setup, Surgery, or End Case. These states are selected by pressing the
associated tab in the menu bar.
Setup
System setup can be accomplished using the six setup panels shown in Figure 2-30,
the Detailed Setup screens, or a combination of both. The Detailed Setup screens
are explained later in this section and include much more detailed information than
the Setup panels. In both the Setup panels and Detailed Setup screens the user can
perform the following actions:
• View available surgical tools and consumables options.
• Connect the surgical tools and options essential for the upcoming operation.
• Prime/Test the surgical tools before commencing surgery.
Basic Setup
The basic setup uses the six setup panels shown in Figure 2-30. These panels are
always displayed in the system state area during setup. Experienced Constellation®
Vision System users will normally use these panels to quickly setup the system. If
desired, the detailed setup screens can be displayed by pressing the Setup button in
the Menu Bar or pressing the expand button on any of the six panels.
Note: Using the barcode scanner to scan a pak automatically sets up these panels
for the accessories in that pak.
The six panels displayed on the Setup Screen are:
• Fluidics Accessories
• Probe • Illuminators
• Handpiece • Lasers