User's Manual Part 5

3.14 8065751025
Constellation
®
Vision System
Extrusion Setup Procedure
The extrusion handpiece connects to the suction/aspiration line from the cassette.
1. Present a sterilized extrusion handpiece to Scrub Nurse.
Circulating Nurse
2. For Combined procedure cassettes: Connect the handpiece to the
aspiration line (light blue) of the irrigation/aspiration tubing set. The
irrigation line (light green) is not used. Use the white male/male luer to
adapt the extrusion handpiece to the blue connector of the aspiration line.
For Posterior procedure cassettes: Connect the handpiece to the auxiliary
aspiration tubing (light blue). Adaptor is not required.
Scrub Nurse
3. Place the handpiece in a sterile cup in preparation for priming. When
aspiration prime is selected, fill the sterile cup with sterile water or BSS.
Ensure the suction port is submerged in the fluid.
Circulating Nurse
4. Press
Start Prime to complete the set up process. Scrub Nurse
Fiber Optic Illuminator Setup Procedure
1. Remove the fiber optic illuminator probe/fiber from the pak, carefully
uncoil the fiber, and present the metal connector to Circulating Nurse.
Scrub Nurse
2. Accept the metal connector from Scrub Nurse and plug it into the desired
illuminator connector. Verify it is fully seated in the adaptor.
Circulating Nurse
3. Slide the protective sheath from the tip of the illuminator probe. Press the
appropriate global Illuminator key to turn the illuminator ON. Adjust the
intensity as needed. Turn illuminator OFF when not in use.
Scrub Nurse
Optional fiber optic illuminator probes can be connected to the second
illuminator connector, supplementing the standard probe. Optional probes
include the following: bare-end, straight pick, pre-bent pick, wide angle,
shielded bullet, illuminated laser, and end irrigating. Some of these
probes deliver substantially less light output than the standard probe,
necessitating higher brightness levels.
Diathermy Setup Procedure
1. Scan the bar code on the Diathermy Kit with the Bar Code Scanner in
the back of the Constellation
®
console. Verify on the set up screen that
Diathermy is indicated on the screen and diathermy ports are illuminated.
Circulator or Non-sterile
Scrub Nurse
2. Present Diathermy Kit to Scrub Nurse.
Circulating Nurse
3. Remove diathermy handpiece from packaging and present connectors to
Circulating Nurse.
Scrub Nurse
4. Accept connectors from Scrub Nurse and connect to the illuminated
Diathermy receptacles
Circulating Nurse
5. Adjust the Diathermy power level as required.
Circulating Nurse