INTRODUCTION
This section of the manual is designed to inform the operator of basic
care and
maintenance of the instrument. If a problem occurs on the instrument, call the Alcon
Technical Services Department and give details of the breakdown circumstances and
effects. From these elements, a specialized technician will evaluate the problem and
determine the maintenance requirements.
For optimum performance, it is the users responsibility to schedule
preventive
maintenance service on the system and its accessories at least one time each year.
Alcon’s Field Service Engineers are trained and equipped to provide the highest
quality of workmanship.
CAUTION
There are no operator replaceable parts other than the fuse. Contact Alcon
Technical Services for all servicing issues.
WARNING!
The Constellation
®
Vision System battery can only be serviced by a factory-trained
Alcon service personnel. Access by untrained personnel can lead to injury.
CARE AND CLEANING
The following tips are recommended for proper care of the
Constellation
®
Vision
System:
Follow
cleaning and maintenance schedules outlined in this section of the manual.
Periodically check chassis appearance.
Pay attention to correct operation of controls, connectors, and indicators.
Damaged hardware must be replaced to ensure safe operation. Call Alcon
Technical Services for assistance.
WARNING!
A qualified technician must perform a visual inspection of the following components
every twelve months:
• Warning Labels (see section one of this manual)
• Power Cord
• Fuses
In case of a deficiency, do not use the system; call Alcon Technical Services.
A qualified technician must check ground continuity for leakage current every
twelve months to ensure they are within the applicable standards (for example:
EN60601-1/IEC601-1). Values must be recorded, and if they are above the applicable
standards, do not use the system; call Alcon Technical Services.