STERILIZATION INSTRUCTIONS
Please consult the accompanying Directions For Use (DFU) for cleaning and sterilization
instructions for Alcon approved reusable accessories. The DFU will provide the recommended
time and temperature guidelines for steam autoclave cycles performed by Alcon, Inc. The sterility
assurance level achieved with these parameters must be validated by each surgical facility.
Please refer to Association for the Advancement of Medical Instrumentation (AAMI) Standards or
your facility’s standard procedures for the most current specifications.
Additionally, per the Sterilizer Equipment Manual, the sterilizer reservoir is to be filled with distilled
or deionized water.
NOTE: The reusable items will withstand steam autoclave cycles at 134° C (273°F). Due to
the variations found in steam autoclaves and the variable bioburden on devices
in clinical use, it is not possible for Alcon to provide specific parameters to ensure
an adequate sterility assurance level. Validation of the individual autoclave, and
verification of the sterility assurance level achieved with a given steam sterilization
cycle, must be performed by each surgical facility. Please refer to below AAMI
Standards or your facility’s standard procedures for the most current specifications.
Please refer to appropriate accessory DFU for proper cleaning instructions.
AAMI TIR No. 12-1994, Designing, Testing, and Labeling Reusable Medical Devices
for Reprocessing in Health Care Facilities: A Guide for Device Manufacturers.
ANSI/AAMI Standards and Recommended Practices Volume 1.1: Sterilization
Designation ST37-1996 (Good Hospital Practice: Flash sterilization – steam
sterilization of patient care items for immediate use)
Designation ST46-1993 (Good Hospital Practice: Steam sterilization and sterility
assurance)