User's Manual Part 5

4.6 8065751025
Constellation
®
Vision System
3.5 Using the Auto Wash Kit, secure the handpiece to the wire mesh basket using
the small gauge wire and connect the handpiece with the “Y” adapter assembly
as shown.
Figure 4-1. Auto Wash: Handpiece with "Y" Adapter Connected
3.6 Place wire basket with handpiece in multi-purpose injector rack and connect the
“Y” adapter assembly to the 4 mm diameter injector nozzle as shown.
Figure 4-2. Auto Wash: "Y" Adapter Connected to Injector Nozzel
3.7 Plug off any unused injector nozzles with silicone tubing.
Figure 4-3. Auto Wash: Unused Injectors Shown Plugged Off
Pictured: Miele* Labwasher, Model G7735 with injector Model #0-177