User's Manual Part 5

5.2 8065751025
Constellation
®
Vision System
System Error Messages
System Error Messages are displayed in a dialog box as shown in Figure 5-2 System
Errors require a partial system shutdown. When an error is detected, the following
actions occur.
Surgical modes related to the error are disabled.
A System Error dialog box is displayed. If the error does not affect the current
active modes, they will remain activated, but the System Error dialog will
still be displayed. The dialog will be removed when the user presses a key to
acknowledge the error.
Activate an error tone.
Keys of unavailable modes will be disabled\grayed. If an unavailable mode key is
pressed, an invalid key tone is emitted and the System Error dialog will again be
displayed.
(TBD) Once a System Error occurs, its status remains until the system power is
cycled. No attempt is made to recover; i.e. modules are not reset and then tested
again. The unavailable modes will not be made available. The exception to this
case is the LPA low-pressure error due to the case of the user triggering this error
during routine setup of the machine. This error can be cleared by correcting the
error condition or by operator acceptance of the condition.
Figure 5-2 System Error Dialog Box