User's Manual Part 5

3.4 8065750XXX
Constellation
®
Vision System
INITIAL SYSTEM SETUP
1. Matching the red dot on the footswitch cable connector to the
red dot on the footswitch, plug cable into footswitch. Plug the
other end of the footswitch cable into the rear panel on the
console (match red dots for proper orientation).
Circulating Nurse
2. Attach air hose to connector on rear panel.
Circulating Nurse
3. Plug main power cord into a suitable wall outlet or receptacle.
Turn Power switch ON located at the bottom of the rear
panel next to the power cord (this switch remains ON in the
I position). Turn system power ON using the Standby switch
located at the middle of the rear panel.
Circulating Nurse
CAUTION
Do not use portable socket outlets with this system.
3. The Setup screen appears if this is the first use of the instrument
after power up. Press the Doctor button and select an available
doctor, or add a doctor by following the steps presented on the
display. Scan the pak or select the appropriate handpiece, tip,
accessories, and procedure types.
Circulating Nurse
4. Sterilize the instruments according to hospital procedure.
Circulating Nurse
CAUTION
U/S and fragmentation handpieces must be at room
temperature before use. Allow handpiece to air cool after
steam autoclave (at least 15 minutes). Never immerse in
liquid to cool.
Circulating Nurse