User's Manual Part 5

3.6 8065751025
Constellation
®
Vision System
The following instructions are applicable to all three procedure paks unless otherwise
specified. For the Combined Procedure Pak setup, follow both Posterior and Anterior
Segment Priming and Setup procedures. For additional assistance in the setup of the console,
press the Setup button on the touch screen monitor.
Procedure Pak and Cassette Setup Procedure
1. Scan the bar code on the Procedure Pak and accessories to be used
with the Bar Code Scanner on the back of the Constellation
®
console.
Verify on the set up screen that the appropriate active accessories are
indicated and selected on the screen.
Circulator or Non-sterile
Scrub Nurse
2. Peel the lid from the pak and aseptically transfer the components to
the sterile field.
Circulator
3. Drape the monitor with the cover provided. Drape the
Constellation
®
tray arm with the Tray arm drape if applicable.
Scrub Nurse
4. Install the cassette by inserting the bottom of the cassette into the
fluidics module. Push the cassette via the handle into the module.
Ensure the drain bag hangs freely.
Scrub Nurse
5. Uncoil the Administration tubing sets (Tubing with Red and Grey
Stripe) and connect to their respective ports on the cassette. The
ports are color coded and keyed for ease of identification. Insert
the connectors into the cassette port and turn clockwise until fully
engaged. For further details refer to the on-screen help.
Scrub Nurse
6. Shake the Drip Chamber on the Administration line and ensure it
rattles. Pass the Administration line to the circulator directly or set the
Administration line on the bottle hook for retrieval by the circulator.
Scrub Nurse
7. Remove the protective cover from the drip chamber and aseptically
spike the irrigation bottle with the drip chamber spike. Squeeze drip
chamber to fill approximately 2/3 to 3/4 full.
Circulator
8. For Constellation
®
Vitrectomy Pak and Constellation
®
Combined
Procedure Pak, proceed to the Posterior Segment Setup Procedure
section. For Constellation
®
Phaco Pak, proceed directly to the
Anterior Segment Setup Procedure Section.
Scrub Nurse