User guide
USE of Alicat instruments (M, MW, MS, MC, MCW, MCS, MCR, MCRW, MCRS, P, PS, PC,
PCS, PCR and PCRS product families only) in Class 1 Division 2 applications.
CSA certies the use of this product for general use as well as use in hazardous
locations as dened by Class 1 Division 2 Group A, B, C and D T4.
CSA certication is indicated by the product label as shown below and
not by the statements in this, or any accompanying documentation.
Special Conditions:
To comply with CSA certication the following information is included in the product literature:
• When equipment is properly labeled, it is suitable in Class I, Division 2, Group A, B,
C and D, T4
o Tamb. -40°C to +50°C
• Electrical Rating 24Vdc, 0.800A max
• Instruments shall be powered by a CSA certied, UL listed, Class II external power
supply suitable for the application
• Instruments shall be housed in an enclosure with a minimum IP54 rating or location
providing equivalent protection
• Instrument’s nal approval shall be provided by the local authority having jurisdiction
EEx nA IIC T4
Class I, Div. 2 Group A, B, C and D T4
24 Vdc, 0.800A max
Class I, Zone 2 AEx nA IIC T4
X – See manual for special conditions
WARNINGS:
EXPLOSION HAZARD – DO NOT
DISCONNECT WHILE CIRCUIT IS LIVE
UNLESS AREA IS KNOWN TO BE NON-
HAZARDOUS.
EXPLOSION HAZARD – SUBSTITUTION
OF COMPONENTS MAY IMPAIR
SUITABILITY FOR CLASS I, DIVISION 2.
www.AlicatScientic.com
Alicat Scientic, Inc.
Tucson, AZ USA
Tel: 520-290-6060
II 3 G
USE of Alicat instruments (M, MW, MS, MC, MCW, MCS, MCR, MCRW, MCRS, P, PS, PC,
PCS, PCR and PCRS product families only) in applications requiring ATEX Certication.
Properly labeled Alicat instruments comply to the following ATEX standard:
II 3 G EEx nA IIC T4 (-40°C ≤ Ta ≤ +50°C)
The examination certicate was issued by the CSA in accordance with
accepted practices and procedures. This conrms compliance with the
European ATEX Directive or Group II Category 3G equipment.
ATEX certication is indicated by the product label as shown above and not by the
statements in this, or any accompanying documentation.
Special Conditions:
• Properly labeled equipment is only certied for use in ambient temperatures in the
range of -40°C to +50°C only
• Electrical Rating 24Vdc, 0.800A max
• Instruments shall be powered by a CSA certied, UL listed, Class II external power
supply suitable for the application
• Instruments shall be housed in an enclosure with a minimum IP54 rating or location
providing equivalent protection
• Instrument’s nal approval shall be provided by the local authority having jurisdiction