USE of Alicat instruments (M, MW, MS, MC, MCW, MCS, MCR, MCRW, MCRS, P, PS, PC, 
PCS, PCR and PCRS product families only) in Class 1 Division 2 applications.
CSA certies the use of this product for general use as well as use in hazardous 
locations as dened by Class 1 Division 2 Group A, B, C and D T4. 
 CSA certication is indicated by the product label as shown below and 
not by the statements in this, or any accompanying documentation.
Special Conditions:
To comply with CSA certication the following information is included in the product literature:
• When equipment is properly labeled, it is suitable in Class I, Division 2, Group A, B, 
C and D, T4
o  Tamb. -40°C to +50°C 
• Electrical Rating 24Vdc, 0.800A max
• Instruments shall be powered by a CSA certied, UL  listed, Class II external power 
supply suitable for the application
• Instruments shall be housed in an enclosure with a minimum IP54 rating or location 
providing equivalent protection
• Instrument’s nal approval shall be provided by the local authority having jurisdiction
EEx nA IIC T4 
Class I, Div. 2 Group A, B, C and D T4
24 Vdc, 0.800A max
Class I, Zone 2 AEx nA IIC T4
X – See manual for special conditions
WARNINGS: 
EXPLOSION HAZARD – DO NOT 
DISCONNECT WHILE CIRCUIT IS LIVE 
UNLESS AREA IS KNOWN TO BE NON-
HAZARDOUS.
EXPLOSION HAZARD – SUBSTITUTION 
OF COMPONENTS MAY IMPAIR 
SUITABILITY FOR CLASS I, DIVISION 2.
www.AlicatScientic.com
Alicat Scientic, Inc.
Tucson, AZ USA
Tel: 520-290-6060
II 3 G
USE of Alicat instruments (M, MW, MS, MC, MCW, MCS, MCR, MCRW, MCRS, P, PS, PC, 
PCS, PCR and PCRS product families only) in applications requiring ATEX Certication.
Properly labeled Alicat instruments comply to the following ATEX standard:
II 3 G EEx nA IIC T4 (-40°C ≤ Ta ≤ +50°C)
The examination certicate was issued  by the CSA in accordance with 
accepted  practices  and  procedures.  This  conrms  compliance  with  the 
European ATEX Directive or Group II Category 3G equipment.
ATEX certication is indicated by the product label as shown above and not by the 
statements in this, or any accompanying documentation.
Special Conditions:
• Properly  labeled  equipment  is  only  certied  for  use  in  ambient  temperatures  in  the 
range of -40°C to +50°C only
• Electrical Rating 24Vdc, 0.800A max
• Instruments shall be powered by a CSA certied, UL  listed, Class II external power 
supply suitable for the application
• Instruments shall be housed in an enclosure with a minimum IP54 rating or location 
providing equivalent protection
• Instrument’s nal approval shall be provided by the local authority having jurisdiction