Installation manual

AT-1500 Plug and Play
30
Experiment with Possible Solutions
This chapter provides the common factors to check in
troubleshooting. It is by no means exhaustive since there are too
many environmental factors to take into account.
The first rule of troubleshooting is to isolate the problem. As you
experiment, vary only one factor at a time. Substitute known
good equipment and see if the problem persists or is eliminated.
Do not overlook the obvious—make sure the card is seated
in the computer and that the cables and connectors are
securely attached.
If the LNK indicator does not light with UTP or fiber
media, make sure the cabling is intact and connected to a
functioning network port.
If the LNK indicator lights on a 10BASE-T segment, but
data transfer is slow, collision-prone, or non-existent, verify
that your cable is of level 3, 4, or 5 (see Chapter 7, Important
10BASE-T Cable Considerations). Level 1 and 2 voice-
quality cable will not work properly.
If the cable is more or less round in section, you probably
have data grade cable, correctly paired, and no transmission
problems. If, however, the cable appears flat in section (as
“Silver Satin” telephone-type cable) you probably have
voice-grade cable and transmission problems. If it looks like
telephone cable, double-check it.
If the computer is on a 10BASE2 segment, make sure the
cable is connected with T-connectors, terminated at the
ends with 50
terminators, and grounded in only one place.
If you suspect the network adapter card is not working
properly, disconnect the computer from the network and
run CardAssistant diagnostics after rebooting with no
drivers loaded.