Specifications

The memory pages are filled with data starting at the address 0. The highest
valid address depends on the number of correction data sets stored in the cam-
era, however it can not exceed 31 (0x1F). If e.g. 27 correction data sets are
available this results in a maximum valid page address of 0x1A.
Both parameters, A and B together, affect the choice and the number of data
sets to be loaded.
By means of the parameter S the address of the active page can be adjusted:
Example
Activate the eleventh correction data set (address 10 = 0xA)
>S=A[CR]
Temperature warning (T)
The command T=1 displays the content of the temperature warning register.
The lowest order bit has the same function as LED L3 placed at the backside of
the camera housing. If the value 1 is output, the temperature of the InGaAs sen-
sor is outside the optimum range.
When you switch on the cold camera, wait some minutes. The sensor tempera-
ture needs this time to stabilize to the set value.
If the temperature warning remains active for a longer period, check the ambi-
ent temperature of the camera.
GigE feature name (CameraSpecialFeatures) Feature Visibility Description
CorrectionDataSet Beginner Number of the correction data
set to activate.
(S=<value> command)
Table 29: GigE feature: CorrectionDataSet
GigE feature name (CameraSpecialFeatures) Feature visibility Description
SensorTemperatureState Expert Camera sensor temperature
state.
0 = The sensor temperature is
OK.
1 = The sensor temperature is
outside the optimum range.
SensorTemperatureStateReg Expert Camera sensor temperature
state register.
QuerySensorTemperatureState Beginner Query camera sensor
temperature state.
(T=1 command)
Table 30: GigE feature: Camera sensor temperature