Specifications

Specifications
Pearleye Technical Manual V2.3.0
16
Pearleye P-030 LWIR
Note
The warranty becomes void in case of unauthorized tampering
or any modifications not approved by Allied Vision Technolo-
gies.
Caution
Surge
To avoid damage caused by surge, connect the camera to an
AC/DC power supply. Use a certified industrial power supply
that complies with common industrial standards. Make sure
the polarization of the power supply is correct.
During the camera start-up, inrush currents ≥ 4 A can occur for
20 ms. Use a sufficiently dimensioned power supply to avoid
damage to the camera.
For the DC (direct current) signal, use cable lengths < 30 m.
Consider that the voltage drop increases with the cable length.
AVT (or your local dealer) provides suitable power supplies:
http://www.alliedvisiontec.com/emea/products/
accessories.html
Feature Specification
Sensor Amorphous silicon uncooled microbolometer focal plane array (FPA)
ULIS UL 04 17 1; built-in electromechanical shutter
Effective chip size 16 mm (H) x 12 mm (V)
Cell size 25 μm x 25 μm
Resolution (max.) 640 (H) x 480 (V)
Lens mount M65 x 0.5
Field of view (FOV) With 18 mm lens: 47.9° x 36.9°
Focus range 0.5 m to infinity (manually adjustable)
Spectral response 8 μm to 14 μm (LWIR)
Pixel format Mono14
Frame rate up to 24 fps
Measurement temperature
range
-20 °C to +80 °C (standard version)
Sensitivity (NETD) typical 80 mK @ 303 K @ f/1.0
Table 4: Specification Pearleye P-030 LWIR