Specifications

Advanced parameters and commands
Pearleye Technical Manual V2.3.0
46
Advanced parameters and commands
This chapter describes the advanced configuration of the Pearleye P-007 LWIR
/ Pearleye P-030 LWIR models.
Two-point correction (A, B, E, J, K, N)
To configure the two-point correction, use the following parameters:
The parameter E controls the operation mode of the two-point correction. The
following value assignments are valid:
Note
In most cases the factory setup of the camera is sufficient.
Only in special cases you need the advanced parameters and
commands.
Parameter Description
A File number of the first correction image recorded at cold
reference temperature.
B File number of the first correction image recorded at warm
reference temperature.
E Two-point correction operating mode
J Set value of the correction image of the actually chosen
correction data set recorded at cold reference temperature.
K Set value of the correction image of the actually chosen
correction data set recorded at warm reference temperature.
N File number for the set values of all correction data sets.
Table 17: Advanced parameters: two-point correction
Value Description
0 Deactivated
Data is passed through transparently.
1 Activated
Correction data A and B with the set values J and K of the
actual correction memory page S is used.
Table 18: Parameter E: values