Specifications

Advanced parameters and commands
Pearleye Technical Manual V2.3.0
60
Example
Set the factory default (115200 Baud and activate serial interface via Gigabit
Ethernet).
>s=2A[CR]
s0
Operation Mode
Integrator / Image
Store
FEDCBA9876543210
--------eaa-xxxx
rw rw rw rw rw rw rw
Bit group Value (hex) Description
x
O110 Baud
1300 Baud
2600 Baud
3 1200 Baud
4 2400 Baud
5 4800 Baud
6 9600 Baud
7 19200 Baud
8 38400 Baud
9 57600 Baud
A 115200 Baud
a
00 RS232 via 15 pin SUB-D only.
01 Additionally activate second channel via Pleora device port serial 0.
10 Additionally activate second channel vis Pleora device port bulk 0
(MODE:UART).
11 Reserved.
e
0 Each character received from the host is echoed back to it.
1No Echo.
Table 31: Baud rates: values