112
After the lens has been focused again the image below will be seen, but now 
with a considerably more uniform gradient.
Note
L
Configuration
To configure this feature in an advanced register: See Table 
125: Advanced register: Shading on page 267.
Note
L
• The SHDG_CTRL register should not be queried at very 
short intervals. This is because each query delays the 
generation of the shading image. An optimal interval 
time is 500 ms.
Note
L
• The calculation of shading data is always carried out at 
the current resolution setting. If the AOI is later larger 
than the window in which correction data was calcu-
lated, none of the pixels lying outside are corrected.
• For Format_7 mode, it is advisable to generate the 
shading image in the largest displayable frame format. 
This ensures that any smaller AOIs are completely cov-
ered by the shading correction.
• The automatic generation of shading data can also be 
enabled when image capture is already running. The 
camera then pauses the running image capture for the 
time needed for generation and resumes after genera-
tion is completed.
• Shading correction can be combined with the image 
mirror and gamma functionality.
• Changing binning modes involves the generation of 
new shading reference images due to a change in the 
image size.