User guide
Configuration of the camera
STINGRAY Technical Manual V4.4.4
291
Temperature register
The temperature is implemented with Presence_Inq=1 (available) and
ON_OFF [6] always ON according to IIDC V1.31:
From -10 °C to +65 °C the temperature accuracy is: +1.5 °C / -2.0 °C
Register Name Field Bit Description
0xF0F0082C TEMPERATURE Presence_Inq [0] Presence of this feature:
0: N/A
1: Available
Abs_Control [1] Absolute value control
O: Control with value in the value field
1: Control with value in the absolute value
CSR
If this bit=1 the value in the value field has
to be ignored.
--- [2..5] Reserved
ON_OFF [6] Write: ON or OFF this feature
Always 1
0: OFF
1: ON
--- [7..19] Reserved
Value [20..31] Temperature at the present time (read only)
Read out temperature value and divide by
10: this is the temperature at sensor in
degree Celsius.
Info: 50 °C at sensor is approximately 45 °C
at camera housing. So never run the camera
with more than 50 °C at sensor.
• min. displayed temperature: -55 °C
• max. displayed temperature: 150 °C
• Increment: 0.25 °C/step
Table 126: CSR: Temperature