Chapter 7: Maintenance
11. Temperature limit testing
a) Decrease the temperature limit threshold from 0.7 V to 0.3 V.
For more details, see Temperature limit threshold on page 89.
b) Increase the temperature reference to 35
C.
When the threshold is overpassed, the red LED (1) in front panel should
light up.
c) Wait the real temperature drops 10
C below the threshold.
d) Reset the instrument with the RESET button (2).
The red LED (1) should turn off.
e) Reset the temperature limit threshold to 0.7 V. For more details, see Tem-
perature limit threshold on page 89.
12. Monitoring: check the output voltage on the rear instrument BNC connector (5).
For 35
C, the voltage should be 350 mV.
13. External reference:
a) connect an external power supply to the External Reference BNC con-
nector (6).
b) Set the voltage to 3.5 V.
Set the switch External/Internal (3) to External.
Set the selector (4) to mode Setting
C.
The temperature displayed (2) should be 35
C.
Set the selector (4) to mode Real
C.
The temperature displayed (2) should follow the reference temperature.
92 QCL user’s manual v3.0