7.2 Calibration procedures
7.2.3. TPG calibration
This procedure describes the steps to be followed to calibrate the TPG front panel
knobs settings. Refer to figure 5.2 for the TPG commands description.
IMPORTANT:
The TPG is calibrated in factory. This procedure is not a routine
procedure and must be performed by experts only.
DANGER
Contact with AC line potential can cause death or serious injury If
work on live electrical circuits is required, use all appropriate
precautions and hot work procedures required at your location.
[1] Verify the +5 V power supply.
[2] During the first test, set all knobs and adjustable capacitors to their medium
values.
[3] Adjust to the knob 50 ns adj (3) in order to obtain a 500 ns long pulse on
TP2 (2).
[4] Measure the pulse period through the boxTrigOut output connector using a
scope.