Datasheet
5.2 TTL Pulse Generator (TPG)
5.2.2. Setting the pulse parameters
1. remove the TPG from the Starter Kit and connect it to the mains independently
2. set the TPG mode to internal
3. connect the TPG to a scope using the Output BNC cable: a 5V TTL signal
should be displayed
4. set the pulse parameters as specified in the datasheet (see section 4.2):
a) set the pulse duration to 50 ns using the 0-200 ns knob (4)
b) set the pulse period to 2.5 µs using the period knob (2) and toggle switch
(3)
5. once the testing conditions are retrieved, set the pulse parameters as desired,
but within a reasonable range of the parameters specified in the datasheet.
Note: the pulse duration should be kept long enough to not distort the TTL
signal.
6. lock the knobs (2) and (4)
7. turn off the TPG and disconnect it from the mains
IMPORTANT:
The operating parameters are specified in the datasheet provided with each laser;
any operation outside the specified range must be agreed with ALPES LASERS or
may result in the loss of warranty.
c