Datasheet

5.3 General procedures
5.3. General procedures
CAUTION:
The laser shall only be operated under conditions as specified in
the datasheet or by ALPES LASERS directly: all other operation may
result in the destruction of the laser and loss of warranty.
5.3.1. QCL startup
In order to insure the laser’s performance and enhance its lifetime, the following steps
must be followed thoroughly:
1. make sure all devices are turned off
2. start cooling for LLH or HHL housing (see section 3.1 for prerequisites)
CAUTION:
The laser must be used at the temperature specified in the datasheet.
Operating the laser below or above the given temperature is likely to
damage the laser and will result in the loss of warranty.
Contact ALPES LASERS or a representant before using the laser at a tempera-
ture outside the range specified in the datasheet.
3. startup the TCU following procedure 5.1.1
4. if the laser is operated in CW mode, follow procedures included in the CW laser
driver OEM manual. Always keep the current 1mA above 0 to avoid reversing
the polarity.
5. if the laser is operated in pulsed mode:
a) startup the external power supply:
i. connect the external power supply to mains
ii. remove the red/black banana cables
iii. turn on the external power supply
iv. set the voltage and current to 0, but keep the voltage 100 mV above 0
to avoid accidental polarity reversal
v. turn off the external power supply
c
2012 SA 65