Datasheet

7.1 Laser installation in LLH housing
7.1.3. Procedure
This procedure allows the installation of a new laser into the LLH housing.
1. disconnect the LLH and open its cover (2), as shown in figure 7.7
2
3
1
4
Figure 7.5.: Exchanging the laser in the LLH housing.
2. remove the two screws (3) holding the contact set (4) using a 2mm hex key
3. remove the contact set using the central PET
1
knob, as show in figure 7.6. Note:
this PET screw is used as a knob, do not remove it.
4. remove the two screws holding the laser submount using a 1.5 mm hex key, as
show in figure 7.6
5. remove the laser submount from its storage box using tweezers
IMPORTANT:
The guidelines presented in section 7.1.1 must be followed carefully.
1
PET refers to polyethylene teraphtalate, a common plastic.
c
2012 SA 79