Process Calibrator TechChek™ 830-KP
CONTENTS GENERAL OPERATING INSTRUCTIONS...........................3 TURN ON ...................................................................3 CONNECTIONS .........................................................3 FIELD & BENCH USE (TILT STAND) ........................4 CHANGING BATTERIES............................................4 CONFIGURING TEMPERATURE SCALES ...............5 ENABLING AUTO-OFF ..............................................5 RESTORING DEFAULT SETTINGS ..........................
BAR, mBar, Atm, Kgf plus torr, inches and mm of mercury or water at a variety of temperatures. Characterized modules for laboratory accuracy in the shop, control room or field. GENERAL TURN-ON Each time you turn on the TechChek 830-KP it runs through a self check then returns to the most recently selected settings. Voltage Calibration Calibrate all your DC millivolt and voltage instrumentation. Source from 0.00 to 110.00 mV and 0.00 to 10.25 V. Read up to 110.00 mV, 11.00 V and 200.0 VDC.
OPERATING INSTRUCTIONS GENERAL FIELD & BENCH USE TechChek 830-KP comes with a carrying case and a built-in tilt stand/hanger. The 830-KP is held securely in the case by VELCRO® for use with the carrying case open. The carrying case also has a snapon belt loop which can be looped around a pipe or rail. The tilt stand is easily raised by pulling the stand until it locks into place. The stand can also be reversed for use as a hanger to suspend the 830-KP.
OPERATING INSTRUCTIONS OPERATING INSTRUCTIONS GENERAL GENERAL SELECTING FUNCTIONS Press a function key to choose among mA, V, T/C, Ω, RTD, FREQ and PRESSURE. DISPLAY CONTRAST The contrast of the Liquid Crystal Display may adjusted for best readability. 1) Press and hold the DISPLAY/SOURCE/READ push-button while turning the 830-KP on to adjust the contrast of the display. 2) Press any UP or DOWN key until the display is most legible.
OPERATING INSTRUCTIONS OPERATING INSTRUCTIONS SOURCE MODE Select source by pressing the DISPLAY/SOURCE/READ pushbutton until the word SOURCE appears on the LCD display. To change the output value, press the UP or DOWN key for each digit. To ramp the output press and hold any UP or DOWN pushbutton. The display will continue to change in increments corresponding to the digit being changed and will automatically carry up and down until the limits of the range are reached.
OPERATING INSTRUCTIONS OPERATING INSTRUCTIONS SIMULATE 2-WIRE TRANSMITTERS 2-WIRE mA, 2-WIRE % (Percent of 4 to 20 mA) 2-WIRE DP % Choose this function to simulate a 2-Wire Transmitter output from 1.00 to 24.00 milliamps. Operates in loops with power supply voltages from 3 to 45 VDC. CALIBRATE VOLTAGE INPUTS V, mV Choose this function to provide an output from 0.00 mV to 110.00 mV and from 0.00 to 10.25 VDC. Current compliance up to 20 mA to provide the driving power to your voltage receivers.
OPERATING INSTRUCTIONS OPERATING INSTRUCTIONS CHECK 1-5 VOLT INPUTS WITHOUT DISCONNECTING WIRES Most 1-5 Volt receivers in 4-20mA loops have a 250 Ohm resistor across the input of the receiver. This resistor may be mounted internally or externally. TechChek 830-KP is connected directly across the input of the 1-5 Volt receiver without disconnecting any field wiring. This saves a great deal of time when a large number of voltage receivers, such as chart recorders or computer systems, require calibration.
OPERATING INSTRUCTIONS OPERATING INSTRUCTIONS CALIBRATE THERMOCOUPLE INPUTS Choose this function to simulate a thermocouple signal into any instrument requiring a thermocouple input. The output of the 830KP is automatically cold junction compensated. CALIBRATE RESISTANCE INPUTS Choose this function to simulate a resistance into a variety of instruments. 1) Disconnect the thermocouple from the instrument being calibrated.
OPERATING INSTRUCTIONS OPERATING INSTRUCTIONS CALIBRATE RTD INPUTS Choose this function to simulate a temperature signal into any instrument requiring an RTD input. CALIBRATE RTD INPUTS 1) Disconnect the RTD from the instrument being calibrated. 2) Press the T/C/RTD/Ω pushbutton until any RTD type (Pt, Ni or Cu) appears on the display. 3) Press the DISPLAY/SOURCE/READ push-button until SOURCE and any RTD TYPE appear on the display 4) Connect using 2 or 3 wires as in the diagrams on the opposite page.
OPERATING INSTRUCTIONS OPERATING INSTRUCTIONS CALIBRATE FREQUENCY INPUTS Choose this function to provide pulses into frequency measuring instruments. The 830-KP output is a zero crossing square wave from -1V to +5V amplitude. Available ranges are from 0.01 to 10.00 kHz, 1 to 1000 Hz and from 1 to 1000 CPM (Counts-PerMinute). CPM is used to simulate extremely slow frequency signals with greater resolution. For example, 10 Hz is equivalent to 600 CPM. To convert from CPM to Hz Divide by 60.
OPERATING INSTRUCTIONS READ MILLIAMP OUTPUTS mA, mA % (Percent of 4 to 20 mA), DP% (DP Flow) Choose this function to measure from 0.00 to +24.00 milliamps. 1) Open the current loop at any convenient point along the signal path 2) Press the MILLIAMP/V push-button until mA or mA and % appear on the display 3) Press the DISPLAY/SOURCE/READ push-button until READ appears on the display 4) Press the mA/%/% DP FLOW push-button to display mA, % 4-20 or % DP Flow.
OPERATING INSTRUCTIONS OPERATING INSTRUCTIONS READ AC VOLTAGES Choose this function to measure from 0.0 to 250.0 V True RMS. MEASURE THERMOCOUPLE SENSORS Choose this function to read a thermocouple. The input of the TechChek 830-KP is automatically cold junction compensated.
OPERATING INSTRUCTIONS OPERATING INSTRUCTIONS READ RESISTANCE CHECK CONTINUITY Choose this function to check continuity. A tone will sound and a sound symbol will appear on the display when the resistance between the leads is less than 100 Ohms. 1) Plug the leads into the TechChek 830-KP as shown below. 2) Press the T/C/RTD/Ω pushbutton until the Ω symbol appears on the display.
OPERATING INSTRUCTIONS OPERATING INSTRUCTIONS MEASURE RTD SENSORS Choose this function to read an RTD. Three wires must be use for both 2 and three wire RTDs. MEASURE RTD SENSORS 1) Disconnect the RTD from the instrument being calibrated. 2) Press the T/C/RTD/Ω pushbutton until any RTD type (Pt, Ni or Cu) appears on the display. 3) Press the DISPLAY/SOURCE/READ push-button until SOURCE and any RTD TYPE appear on the display 4) Connect using 3 wires as in the diagrams on the opposite page.
OPERATING INSTRUCTIONS OPERATING INSTRUCTIONS COUNT FREQUENCIES Choose this function to use the 830-KP as a frequency counter. Available ranges are from 0.01 to 15.00 kHz, 1 to 1500 Hz and from 1 to 1500 CPM (Counts-Per-Minute). To measure waveforms with amplitudes between 1 V and 10.25 V RMS use the low level inputs. Use the high voltage connection to read waveforms with amplitudes from 10.25 to 250.
OPERATING INSTRUCTIONS OPERATING INSTRUCTIONS CALIBRATE TRANSMITTERS ANY SOURCE FUNCTION AND READ mA, READ %, READ DP% Choose this function to supply the input signal to the transmitter and displaying the 4-20 mA output of the transmitter (used with 4Wire Transmitters). ANY SOURCE FUNCTION AND P-XMTR mA, P-XMTR %, P-XMTR DP% Choose this function to simultaneously supply power to a 2-Wire transmitter while supply the input signal to the transmitter and display the 4-20 mA output of the transmitter.
OPERATING INSTRUCTIONS OPERATING INSTRUCTIONS READ PRESSSURE READ PRESSURE Choose this function to measure pressure with a ModPak 91 module holder and one of the many QuikCal 90 pressure modules. WARNING! - SAFETY CONSIDERATIONS Sudden release of compressed or stored gas can cause personal injury. Always vent the system before making pressure connections or disconnections. Exercise standard physical protection practices such as eye protection, gloves, protective clothing, etc.
OPERATING INSTRUCTIONS CALIBRATE PRESSURE TRANSMITTERS READ PRESSURE AND P-XMTR mA, P-XMTR %, P-XMTR DP% Choose this function to simultaneously supply power to a 2-Wire pressure transmitter while supplying the input signal to the transmitter and displaying the 4-20 mA output of the transmitter.
SPECIFICATIONS SPECIFICATIONS GENERAL POWER & MEASURE 2-WIRE TRANSMITTERS RANGES & ACCURACY: Same as for MILLIAMP SOURCE OUTPUT CURRENT: up to 24.00 mA TYPICAL DRIVE CAPABILITY:1200 Ohms @ 20.00 mA COMPLIANCE VOLTAGE: nominal 25 VDC @ 20 mA COMMON MODE ERROR: 0.01% Full Scale/Common Mode Volt TYPICAL 90 DAY ACCURACY: ±(0.025% of Full Scale + 1 LSD)1 1 YEAR ACCURACY: ±(0.05% of Full Scale + 1 LSD) WARM UP TIME: 10 seconds to specified accuracy, 2 minutes to maximum accuracy TEMPERATURE EFFECT: ±0.
SPECIFICATIONS SPECIFICATIONS MEASURE DC VOLTS RANGES: 0.00 to 110.00 mV; 0.00 to 10.25 V; 0.0 to 200.0 V ACCURACY: ±(0.05% of 110 mV+ 0.01mV) = ±0.07 mV ±(0.05% of 10.25 V + 0.01V) = ±0.02V ±(2% of 200.0 V + 0.1V) = ±4.1 V INPUT RESISTANCE: >1 Meg Ohm to 10.25V, >5 Meg Ohm to 200V SOURCE RESISTANCE EFFECT: 0.01% per 100 Ohms SOURCE RTD & OHMS RTD TYPES: Pt 100Ω for 1.3850 (DIN/IEC 751 & New JIS), 1.3902 (Burns), 1.3926 (US Lab) & 1.
THERMOCOUPLE SPECIFICATIONS T/C TYPE °C RANGE J 100 to 1200 -50 to 99 -150 to -49 -200 to -149 K THERMOCOUPLE SPECIFICATIONS °F RANGE ACCURACY ±1.7 ±1.8 ±2.2 ±2.8 212 to 2192 -58 to 211 -238 to -57 -328 to -237 ±2.3 ±2.5 ±3.2 ±4.3 N 300 to 1300 100 to 299 -50 to 99 -200 to -49 1100 to 1372 0 to 1099 -100 to -1 -200 to -99 ±2.2 ±2.0 ±2.3 ±3.6 2012 to 2500 32 to 2011 -148 to 31 -328 to -147 ±3.1 ±2.9 ±3.3 ±5.7 R T 200 to 400 0 to 199 -100 to -1 -200 to -99 ±1.7 ±1.9 ±2.4 ±3.
RTD SPECIFICATIONS QuikCal 90 MODULE SPECIFICATIONS RTD TYPE ALPHA °C RANGE ACCURACY Pt 100Ω (DIN/IEC/JIS 1989) 1.3850 -100 to 850 ±1 OPERATING TEMPERATURE: -10°C TO 50°C (13°F TO 122°F) STORAGE TEMPERATURE: -40°C TO 85°C (-40°F TO 185°F) WEIGHT: 0.4 kg (14.5 oz) TEMPERATURE EFFECT: None (Compensated over full range) CONNECTION: 1/8” NPT FEMALE MEDIA COMPATIBILITY: Any liquid or gas compatible with 316 stainless steel and fluorocarbon rubber Pt 100Ω (Burns) 1.
WARRANTY Altek products are warranted to be free from defects in material and workmanship (excluding fuses, batteries and leads) for a period of three years from the date of shipment. Warranty repairs can be obtained by returning the equipment prepaid to our factory. Products will be replaced, repaired, or adjusted at our option. Altek gives no other warranties, including any implied warranty of fitness for a particular purpose.
PN 1800470 Rev A April 2002 ©2002 Altek Industries, Inc. Specifications subject to change without notice. All rights reserved. Printed in U.S.A.