Specifications
AlterG M300 Medical Operations Manual Part Number 102508 Rev D
13
SAFETY INFORMATION, WARNINGS AND
CAUTIONS
Before using the AlterG please familiarize yourself with this manual so
that you may operate the AlterG in a safe and effective manner.
Instructions may be found throughout this manual, on the control
console and on labels applied to the machine. We want your
experience with the AlterG to be safe and enjoyable, so please make
sure you read and understand this entire manual before operating your
system.
DANGER: Imminently hazardous situation to be avoided
that will result in serious injury or death.
DO NOT modify the plug provided with the treadmill. The AlterG
requires a 20A 220 VAC outlet protected by a 20 amp circuit
breaker. The AlterG comes equipped with a 6-20P NEMA plug.
If the provided plug will not fit in the outlet, have a proper outlet
installed by a qualified electrician.
Do not use any electrical adapters. To do so could result in an
electrical shock hazard.
Do not operate the AlterG in wet or damp environments.
Do not operate the heart rate monitor transmitter in conjunction
with an electrical heart Pacemaker or similar device. The
transmitter may cause electrical disturbances, which can
interfere with pacemaker function.
Always unplug the AlterG before cleaning or servicing.
Do not soak any part of the AlterG with liquid during cleaning;
use a sprayer or damp cloth. Keep all liquids away from electric
components. Always unplug the machine before cleaning and
maintenance.
Only an authorized technician should service the system.
Do not place any liquids on any part of the AlterG.
Do not unplug or alter any of the internal wiring on the machine
after installation.
Do not operate where aerosol (spray) products are being used.
!