SECTION 2: SETUP AND INSTALLATION
An AlterG qualified technician will install the AlterG Anti-Gravity
Treadmill after delivery. Please make sure that you inspect the AlterG
upon delivery for any damage that may have occurred during
transportation. Take pictures and immediately report any damage to
the shipping company and AlterG. When you sign for the shipment of
your AlterG, you are taking responsibility for any damage that may
occur before installation.
ELECTRICAL REQUIREMENT
The recommended electrical outlet/power source for the AlterG Anti-
Gravity Treadmill is a 20 ampere, 220VAC @ 60 Hz dedicated circuit
with ground. The plug supplied with the AlterG Anti-Gravity Treadmill
is designated by the NEMA configuration system as 6-20P. The
corresponding receptacle for the plug is a NEMA 6-20R.
International systems are shipped without a plug on the power cord.
An appropriate plug for the country and facility in question should be
wired as follows:
Black Conductor: Line
Black Conductor: Line
Green/Yellow Conductor: Ground
GROUNDING REQUIREMENTS
The AlterG must be grounded electrically. If there is an electrical
malfunction, grounding provides a path of least resistance for electric
current to reduce the risk of electric shock. The AlterG is equipped
with a power cord with a grounded plug (domestic systems). This plug
must be plugged into an appropriate receptacle (NEMA 6-20R) that is
properly installed and grounded in accordance with the current
National Electrical Code as well as all local codes and ordinances. If
you are at all unsure of these requirements, contact AlterG for
clarification.
LOCATION REQUIREMENTS
The AlterG is 38 inches wide and 84 inches long. It weighs
approximately 750 pounds. It needs to be placed on a structurally
sound surface. If it is used above ground level it is advisable to place it
near the corner of the room or where the floor will be strongest to
ensure maximum support during high-speed, high-impact use. The
surface should be level to ensure minimal flexing of the AlterG frame.
It is recommended that the AlterG be placed on a rubber mat or
padded indoor/outdoor style carpet. The front of the AlterG needs to be
within 10 ft (3.05 meters) of the proper electrical outlet. Check with a