Specifications
AlterG M300 Medical Operations Manual Part Number 102508 Rev D
41
SEMI-ANNUALLY:
Check motor brushes for wear. Dress commutator with stone if
needed.
ANNUALLY:
Note: AlterG recommends that you have a qualified
technician perform the annual maintenance.
1. Perform the following procedures annually:
- Inspect all nuts and bolts. Tighten any that are loose.
- Remove the bag and thoroughly vacuum under and
around the treadmill.
- Clean the running surface, if necessary use a bristle
brush to remove heavy grime.
- Check and adjust treadbelt alignment if needed.
- Wipe down the treadmill deck under the treadbelt.
- Lubricate the treadmill deck with SlipCoat.
- Vacuum under the treadmill motor cover.
- Adjust the drive belt tension. Replace the belt if worn or
damaged.
- Inspect the treadmill motor brushes and replace if worn
below 3/8 inch. Dress the commutator if needed.
SEASONALLY:
1. In autumn & winter the drier climate in many regions of the
country may cause a static charge build-up to occur when the
treadmill is used. Spray the running surface with a staticide
spray to prevent static shock to treadmill users and to prevent
interference with the treadmill’s electronic systems. A worn out
treadbelt can also contribute to a static problem. Examine the
belt for excessive wear and replace if necessary.
CLEANING THE WINDOWS
The windows on your AlterG are made of Strataglass™
6
. Special care
must be taken to ensure they remain clean and clear. Strataglass
recommends the use of IMAR Strataglass Protective Cleaner for
general cleaning. Other cleaning products may dull the clear finish.
You should do this cleaning regularly (about once a week, depending
on usage).
6
Strataglass is a trademark of Strataglass, Inc., Fort Lauderdale, FL