Specifications

AlterG M300 Medical Operations Manual Part Number 102508 Rev D
50
When replacing the brush, make sure it slides with no resistance inside
the holder. If resistance is felt, dress the sides of the brush with emery
cloth or a small file until it slides freely.
Replace the brush holder by pushing it in place until it clicks.
Figure 24. Sliding the Brush Holder in Place
Inspect motor commutator for wear (black scoring present on copper
segments). Dress with emery cloth if scored.