Digital Monitor AMC-1DBx Series User Manual
84
database and self-test will be
performed on data base
storage device, if any test
fails:
FAULT LED will remain
lit with other
combinations of D1-D3
and ALARM1-3 LEDs.
The monitor will reboot
over and over.
These monitor faults
must be reported to
qualified technician for
repair. Do not advance to
the next step.
5. Establish communications with downstream
peripherals; relays, analog outputs and sensors.
If the system detects a
missing device the FAULT
LED will turn on, the external
buzzer and Annuniciator will
also be activated.
6. Establish communications with Upstream peer and
masters.
If the system detects a
duplication or configuration
problem with the IP and/or
BACnet interfaces. The
System LED will turn off, the
extern buzzer and
Annuniciator will also be
activated.
6.13.2
Adding a New Transmitter.
New transmitters will be added to the monitor when customers increase the number of
transmitters, either as a result of increased floor plan space or additional gas types. Follow
the steps outlined in the following table to add a new transmitter:
Step Action Comment
1. Set the monitor’s
appropriate RS-485
Interfaces to “Admin
Disable” mode.
It is very likely that MODBUS errors will be seen
during this install. Disabling the monitor’s
appropriate RS-485 will force the alarming functions
into fail state for that interface. Audio alarms can be
disabled with position 1 of dipswitch.
2. Install the transmitter. The mounting instructions for transmitters are
provided in the transmitter user manual.
Select an appropriate location for the transmitter
with the following considerations: