ADC Pocket Diagnosic Instruments ® Use, Care and Maintenance ®
Thank you for choosing an ADC® Diagnostic set. We’re proud of the care and quality that goes into the manufacture of each and every diagnostic instrument that bears our name. Every component has been carefully designed to maximize performance. This booklet refers to both Otoscope and Ophthalmoscope sets (Models 5110N, 5110E, 5111N, 5112N).
(including internal liners) can NOT be cleaned or disinfected and should be safely discarded if contaminated. Do not carry the case into a contaminated environment. WARNING: When replacing lamp, allow lamp to cool for five minutes before handling. WARNING: No form of sterilization has been validated for this device or its corresponding components and sterilization may compromise the safety and effectiveness of this device. (This includes steam sterilization.
ation. To select, rotate the aperture selection wheel at the front of the instrument head. Corrective Lens Selection This unit is equipped with 19 corrective lenses from +20 to -20 diopters (negative lenses in red). To select, rotate the corrective lens selection wheel at either side of the instrument head.
Disposable Specula Cleaning/Disinfection Instructions Disposable specula should not be cleaned or disinfected in any way. Disposable specula are single patient use only and must be discarded after use. Disposable specula should not be used if they appear to be visibly contaminated or have accidentally come into contact with contaminated materials prior to use. Troubleshooting The following guidelines should be used to determine if your device has reached its end of life or requires servicing.
Ophthalmoscope Troubleshooting Guide Trouble Area Possible Cause Corrective Action Aperture dial is between positions Rotate Aperture dial Lamp Burned out Replace the lamp with ADC part #5112N-4 Wrong Lamp Installed Replace the lamp with ADC part #5112N-4 Batteries are depleted Replace with fresh “AA” batteries Lamp not fully inserted Re-insert lamp Thread/connection is damaged Service required The aperture dial is not centered Move aperture dial to the full detent position Lamp near end of
Do not dispose of this product as unsorted municipal waste. Prepare this product for reuse or separate collection as specified by Directive 2002/96/EC of the European Parliament and the Council of the European Union on Waste Electronic and Electrical Equipment (WEEE). If this product is contaminated, this directive does not apply. LIMITED WARRANTY American Diagnostic Corporation (ADC®) warrants its products against defects in materials and workmanship under normal use and service as follows: 1.