Owner manual

Rev. 5 115
Remote Interface Reference
SETUP Configuration Commands and Queries
ABsorber?
Returns “0” indicating that an energy absorber is not present in the
system, or “1” indicating that an energy absorber is present.
CONFigure:RAMP:RATE:UNITS {0|1}
Sets the preferred ramp rate time units. Sending “0” selects seconds. A “1”
selects minutes. “0” is the default value. The selected units are applied to
both the Model 430 Programmer display and the appropriate remote
commands.
RAMP:RATE:UNITS?
Returns “0” for ramp rates displayed/specified in terms of seconds, or “1”
for minutes.
CONFigure:FIELD:UNITS {0|1}
Sets the preferred field units. Sending “0” selects kilogauss. A “1” selects
tesla. “0” is the default value. The selected field units are applied to both
the Model 430 Programmer display and the applicable remote commands.
FIELD:UNITS?
Returns “0” for field values displayed/specified in terms of kilogauss, or “1”
for tesla.
CONFigure:IPNAME <system name>
Sets the system name (also known as host name or computer name), the
name by which the Model 430 Programmer is identified on a network.
Note
If the system name value is changed, the Model 430 Programmer
power must
be cycled off for at least 5 seconds and then back on to
complete the change. The new value will be used internally
immediately (even before cycling power off and back on), but the
IPNAME? query will return the previous system name until the
Model 430 Programmer is restarted.
IPNAME?
Returns the system name (also known as host name or computer name).
4.5.4 Protection Commands and Queries
The Protection Commands and Queries provide read/write access to all
protection settings which can be accessed under the Settings Protection
submenu. See section 3.10.3.6 for more information regarding the settings
protection features of the Model 430 Programmer.