Problem #3: Enbrel Mini
®
single-dose
prefilled cartridge is defective or not
recognized as an Amgen product.
Action: Remove Enbrel Mini.
Replace with a new Enbrel Mini.
Call 1-888-4ENBREL (1-888-436-2735)
for a replacement Enbrel Mini.
Problem #4: Door is held closed for more
than one minute.
Action: Remove your hand from the door.
When the door opens, remove Enbrel Mini
from AutoTouch Connect
TM
, if present.
Wait for the error symbol to stop blinking
and close the door. If you have not
given yourself an injection, place a new
Enbrel Mini in AutoTouch Connect and
continue. If the error symbol remains on,
call 1-888-4ENBREL (1-888-436-2735).
Problem #5: AutoTouch Connect has
been dropped.
Action: A dropped AutoTouch Connect is
not safe to use. Call 1-888-4ENBREL
(1-888-436-2735) for a replacement.