User's Manual

22
Needle Exposure
Symbol: Orange Enbrel Mini
®
single-dose
prefilled cartridge, a red triangle with an
exclamation point, red status button, and
error sound.
Problem: A problem occurred during an
injection and the needle may be exposed.
Action: If there is still fluid in EnbrelMini,
an incomplete dose may have been
injected. Call your healthcare provider
if you feel you have given yourself an
incomplete injection. Call 1-888-4ENBREL
(1-888-436-2735) for further assistance
with your AutoTouch Connect
TM
.
Use caution when both Enbrel Mini
problem symbol and the AutoTouch
Connect reusable autoinjector failure
symbols are lit as the needle may be
exposed. Take special care when removing
and handling Enbrel Mini. Remove
Enbrel Mini then put it in an FDA-cleared
sharps disposal container.
Troubleshooting:
Error Symbols