Reusable Autoinjector Failure
Symbol: Red triangle with an exclamation
point, red status button, and error sound.
Problem: Several errors have occurred or
AutoTouch Connect has stopped working.
Action: Reset AutoTouch Connect.
To reset AutoTouch Connect:
Hold AutoTouch Connect away from skin
and press the status button to wake
AutoTouch Connect. The failure symbol
should begin blinking and a chime
should sound.
While the failure symbol is blinking, press
and hold door button until all symbols
are temporarily displayed and the status
button blinks green. The door button
should be held for at least 10 seconds.
After a successful reset, if an Enbrel Mini
®
single-dose prefilled cartridge is still inside
AutoTouch Connect, remove it. Close the
AutoTouch Connect door. Then, for the
next injection, start by pressing the door
button to open Enbrel Mini door.
If AutoTouch Connect does not respond
after three attempts to reset, call
1-888-4ENBREL (1-888-436-2735).