User's Manual

25
Injection does not start when
pressing the status button.
Reason #1: Place the AutoTouch
Connect
TM
reusable autoinjector on your
skin and wait for the status button to
turn green. A skin sensor is located
on the injection end. You cannot start
an injection unless the injection end of
AutoTouch Connect is touching your
skin. Hold the injection end on your
skin throughout the entire injection.
Reason #2: AutoTouch Connect may
be asleep. To conserve battery power,
AutoTouch Connect goes into a “sleep
mode” after three minutes of no activity.
If it seems unresponsive/asleep, remove
AutoTouch Connect from your skin and
press the status button to wake it up.
An Enbrel Mini
®
error symbol
appears immediately after loading
the EnbrelMini single-dose
prefilledcartridge.
Reason: This will happen if the purple
cap has been removed before inserting
Enbrel Mini into the door. Do not reuse
or recap. Begin again using a new
Enbrel Mini.
Do not remove the purple cap until
after Enbrel Mini has been inserted into
AutoTouch Connect.