Installation and Maintenance Manual Medical Gas Outlet British Standard Slim Style Compatible w w w . a m i c o .
Table of Contents Product Description Cleaning and Lubrication Inspection and Testing 3 3 4 Installation and Dimensions Console Mount Surface Mount 5 6 Service Preventive Maintenance Latch-Valve Assembly AGSS Flow Control Adjustment 7 8 - 10 11 Model Numbers Latch-Valve Assembly 12 Rough-in Assembly 12 Complete Outlets 12 Gas indexing 13 Replacement Components Gas Latch-Valve Assembly Vacuum Latch-Valve Assembly AGSS Latch-Valve Assembly Rough-In 14 15 16 17 2 Amico Pipeline
Product Description The Amico Medical Gas British Outlet Slim Style is composed of two separated modules: the “Rough-in assembly” and the “Latch-valve assembly”. The “Rough-in assembly” consists of a brass machined body that incorporates a spring loaded check assembly. The copper pipe is sliver brazed into the body and flared to 12 mm inside diameter for external pipeline connections. The brass body and pipe assembly are inserted in to a gas specific plate.
Inspection and Testing Pressure drops across the Amico British Outlet shall comply with clause 6.3 of BS EN 737-1:1998. Nominal Pressure (kPa) Test Pressure (kPa) Test Flow (1/min) EN 737-1 Limit (kPa) 400 320 60 15 400 320 200 70 700 640 300 70 Vacuum *40 40 15 * Absolute pressure AGSS Requirements for flow and pressure drop across terminal units with probe inserted clause 5.4.
Installation and Dimensions Console Mount www.amico.
Installation and Dimensions Surface Mount 6 Amico Pipeline
Preventative Maintenance Regular inspection and maintenance of the Outlets will prolong its life and reduce the possibility of sudden, inconvenient component failures. The use of damage probes or faulty equipment may require further maintenance. Outlets should be subjected to regular inspection and testing as follows: Monthly; a. Visually inspect the Outlet for signs of damage. b. Check to see if the Front Latch Assembly (2nd Fix) operates smoothly with the gas specific probe. c.
Gas Latch Valve Assembly 11 5 9 10 8 4 7 6 2 1 3 1. Unscrew the two Surface Box Cover screws (1), and remove the Surface Box Cover (2), or pry off Trim Ring (3) with a flat screw driver. 2. Unscrew the two Retaining screws (5) until the Latch Valve Assembly (4) can be removed from the Outlet. 3. Remove the two Retaining washer screws (11), and remove the Retaining washer (10). 4. Remove the Gas Connector (8) 5.
Vacuum Latch Valve Assembly 12 11 10 5 9 8 4 7 2 6 1 3 NOTE: Suction Outlets do not have a secondary check valve. Suction source to the Vacuum Outlet must be turned off before removing the Latch-Valve Assembly 1. Unscrew the two Surface Box Cover screws (1), and remove the Surface Box Cover (2), or pry off Trim Ring (3) with a flat screw driver. 2. Unscrew the two Retaining screws (5) until the Latch Valve Assembly (4) can be removed from the Outlet. 3.
AGSS Latch-Valve Assembly 13 12 9 5 7 4 11 8 6 2 1 10 3 1. Unscrew the two Surface box Cover screws (1), and remove the Surface Box Cover (2), or pry off Trim ring (3) with a flat screw driver. 2. Unscrew the two Retaining screws (5) until the Latch Valve Assembly (4) can be removed from the Outlet. 3. After removing the Latch Valve Assembly (4) from the Outlet, you can then pull the Check Valve Body Assembly from the back side of the Latch Valve Assembly (4).
AGSS Flow Control Adjustment Referencing Part Numbers: O-BSQSL-SUR-AGS O-BSQSL-CON-AGS www.amico.com A.G.S.S. Threaded Nose Valve Flow Control Screw (3mm Allen Key) 1. Insert a 3 mm Allen Key into the front of the Check Valve. Surface Mount Box British Standard A.G.S.S. 2. Adjust the Flow Control Screw to desired setting. www.amico.
Rough In Assembly 4 3 1 2 CAUTION: Ensure that the supply pressure is shut off before performing service. Inside the “Rough-in assembly” is a secondary check valve whose function is to shut off gas flow when the “Latch-valve assembly “is removed. This seal O-Ring also prevents leakage around the latch valve connector. As the secondary seal is only a static seal, it will rarely need replacement.
Model Numbers Surface Mounted Latch Valve Assembly Rough-in Assembly Complete Assembly Oxygen O-FASC-BSQSL-OXY O-BAKSUR-BSQSL-OXY O-BSQSL-SUR-OXY Medical Air (400 kPa) O-FASC-BSQSL-MA4 O-BAKSUR-BSQSL-MA4 O-BSQSL-SUR-MA4 Medical Air (700 kPa) O-FASC-BSQSL-MA7 O-BAKSUR-BSQSL-MA7 O-BSQSL-SUR-MA7 Vacuum O-FASC-BSQSL-VAC O-BAKSUR-BSQSL-VAC O-BSQSL-SUR-VAC Nitrous Oxide O-FASC-BSQSL-N20 O-BAKSUR-BSQSL-N20 O-BSQSL-SUR-N20 Anaesthetic Gas Scavenging System O-FASC-BSQSL-AGS O-BAKSUR-BSQSL-AG
Gas Indexing As seen from the front of the “Rough-In Assemblies” VACUUM OXYGEN NITROUS OXIDE MEDICAL AIR 400 kPa MEDICAL AIR 700 kPa 14 Amico Pipeline AGSS
Replacement Components Gas and Vacuum Latch Valve Assembly 12 11 5 10 9 8 4 7 2 6 1 Item 3 Description Model Numbers 1 Surface Box Cover screws (2 req'd) H-STPP-6150AB 2 Surface Box Cover O-X-BSQC-SBX2-375 3 Trim Ring O-X-BSQC-FACTRM2 4 British Standard Slim Style Latch-Valve See Above 5 Outlet Retaining Mounting screws (2 req'd) H-MPP-0608 6 Body Seal O-X-OXE-SEAT* 7 Poppet O-X-BSQC-POPET2* 8 Gas connector O-X-BSQC-VLV2 9 Poppet O-Ring H-ORING105VTN* 10 Retaining wash
Replacement Components AGSS Latch Valve Assembly 13 12 11 5 9 4 7 2 3 Description Model Numbers 1 Surface Box Cover screws (2 req'd) H-STPP-6150AB 2 Surface Box Cover O-X-BSQC-SBX2-375 3 Trim Ring O-X-BSQC-FACTRM2 4 British Standard Slim Style Latch-Valve See Above 5 Outlet Mounting screws (2 req'd) H-MPP-0608 6 AGSS Check Valve O-X-BSQC-CHKVLV 7 Flow Adjustment screw H-HSSS-CPM601-6SS 8 Locking screw H-HSSS-CPM601-6SS 9 AGSS Check Valve Spring O-X-BSQC-VLV-SPR* 10 Check
Replacement Components Rough-in Assembly 4 3 2 1 Item Description Model Numbers 1 Secondary Seal O-Ring H-ORING-115VTN* 2 Secondary Check Valve O-X-BSQC-BKCHK2* 3 Secondary Check Valve Spring O-X-BSQC-BAKSPR* 4 Rough-in Body See above Above parts with an “*” are found in repair kit: O-RK-BAK-BSQSL www.amico.
www.amico.com Amico Pipeline | www.amico.com 85 Fulton Way, Richmond Hill Ontario, L4B 2N4, Canada Toll Free Tel: 1.877.264.2697 Toll Free Fax: 1.866.440.4986 Tel: 905.763.7778 Fax: 905.763.8587 Email: info@amico.