CPU Mount
WARNING Ensure CPU is within weight and size rating of the CPU mount. If not, please contact Amico
Accessories at (1-877-264-2697)
WARNING The CPU mount can be mounted anywhere at any height on the rail. However, DO NOT mount
anywhere above a patient’s head.
WARNING Ensure rail is stable and properly installed before mounting FIT Station.
1. Determine a suitable location on the MRS for the CPU to make sure it isn’t above a patient’s head.
2. Lower the CPU mount down the MRS to the desired height. Tighten the two set screws in the Mounting Adapter
to the tightest possible position using a 1/8” HEX key. (Figure 1)
3. Mount the CPU and secure it by tightening the two thumb knobs.
(Figure 1)
Thumb
Knobs
CPU
Mount
MRS
1/4-20 Set Screws on
Mounting Adapter
CPU slides
into slot
OSHPD Anchor Pre-Approval OPA-1743
In compliance with standards established for the anchor and installation of instrument support systems by the
California Oce of Statewide Health Planning and Development (OSHPD), Amico Accessories Inc. has obtained pre-
approval for MRS 19’ (48.26cm) Wall Channels or longer with extension Arm. For proper installation requirements
please refer to our OPA Document.
SECTION 2: Installation on Rail Systems